NoDec Bari 1200

The NoDec Bari 1200 is an automatic, high performance 3 cell cycle alternating pressure mattress designed to replicate the body’s natural spontaneous movement.  With vacuum assisted cell deflation, which ensures complete evacuation of air from deflating cells, it is clinically demonstrated to provide complete interface pressure elimination at regular Intervals for bariatric patient groups up to and including patients of 400kg. It is recommended for the prevention of pressure injuries in patients at High to Very High Risk of pressure injury development as well as for healing established Stage 3 and 4 pressure injuries.

With a unique combination of pressure elimination and comfort modes the NoDec Bari provides the ultimate protection and comfort for the most vulnerable bariatric individuals and assists in the management of all grades of pressure injuries.

The maximum patient weight is determined by the width of the mattress and NoDec Bari is available in 2 sizes.

NoDec Bari requires a bed platform of 1000mm platform up to a 330kg patient; a 1200mm platform up to a 400kg patient.

Available Sizes:

NB 1000 - 2000 x 1000 x 250 mm / SWL 350kg
NBXL 1200 - 2000 x 1200 x 250 mm / SWL 400kg

 

https://www.traditionrolex.com/16
Vacuum assisted deflationEnhances the pressure elimination phase to aid the treatment of even the most severe pressure ulcers
Solenoid technologyEnhances pressure relief by providing a more efficient flow of air in and out of the cells therefore a superior reactive hyperaemic response in the deep tissues about the bony prominences
Easily accessible CPRIntuitive, always visible and accessible CPR release valve provides rapid deflation in approx. 10 seconds.
Side formersPermanently inflated side formers provide a firm mattress edge to ensure patient safety and support during exits or transferring to a trolley/chair.
Comfort controlAdjustable comfort control to Firm, Medium and Soft settings provide nursing staff with the ability to adjust air cell inflation pressures in either Static or Dynamic mode to suit individual patient comfort or clinical requirements.
Timed Static modeProvides a firm and stable surface for essential nursing procedures such as physiotherapy and wound cleaning. A safety automatic return to Dynamic mode after 30 minutes ensures that patients are not put at risk of pressure damage.
30 hour Transport modeEasy to operate one way valve lets mattress remain inflated for at least 30 hours and also facilitates patient/bed movement without unnecessary risk to patient.
Audible and visual alarmsA sophisticated set of audible and visual alarm functions provide enhanced patient safety and ensures peace of mind for nursing staff.
Dartex cover with welded seams Soft, easy to clean, waterproof Dartex cover provides a complete barrier against viruses and blood, minimising infection control risks.
Range of mattress widths Available in 1000mm/1200mm widths to provide optimum flexibility to cater for range of bariatric patients.
Product CodeNBXL
Risk CategoryVery High Risk
Therapeutic Safe Working LoadExtraWide 1200 : 400kg
Warranty2 years
Mattress Dimensions 1900mm x 1000mm x 250mm (Wide)
1900mm x 1200mm x 250mm (Extra Wide - see NDXL)
Power Unit Dimensions350mm x 180mm x 380mm (LxWxH)
Product WeightMattress: 11kg Power Unit: 6.8kg
Cycle Time7.5 Minutes (1-in-3 cell cycle)
Cycle ControlMicroprocessor controlling solenoid valves
Supply Voltage230V, 50 Hz
Power RatingMax – 100 Watts
Electrical SafetyClass II, Type B
ComplianceEN60601-1, IEC60601-1